Mounting the eyepiece
· The MM6-OCC10X eyepiece with cross hairs should always
CAUTION
be attached to the right microscope sleeve (in case of a
binocular tube).
· Use the MM6-OCC10X eyepiece for the erect image
monocular observation tube MM6-EMO.
1
Loosen the clamping knob b of the right eyepiece sleeve using the
tool provided with the frame and insert the MM6-OCC10X carefully
into the sleeve a .
2
When the cross hairs and the stage travel in the XY direction are adjusted,
tighten the clamping knob b using the tool provided with the frame.
} If you want to use a clamping knob without tab, tighten the
clamping knob provided with MM6-ETR/MM6-EMO using the
dedicated Allen screwdriver.
Do not tighten the clamping knob too firmly.
CAUTION
3
Insert the MM6-OC10X into the left eyepiece sleeve.
At this stage, clamp the eyepiece sleeves temporarily and
CAUTION
lightly. When the microscope is assembled after
completing the setup of the light source, etc., perform the
following adjustments.
Adjusting the cross hairs and X- and Y-travel direction of stage
Place a test bar or a straight edge on the table approximately parallel with
the table travel direction, then adjust the positioning so that the intersection
of the eyepiece cross hairs comes on the straight line of the test bar.
Next, rotate the eyepiece on the parallel straight line to align the cross
hairs, and tighten the clamping knob.
} When the AF unit is combined with the MM6C-KMAS or MM6C-
RLAS, the periphery of the observation field of view may become
dark. If this bothers you, replace the cross-frame part of the eyepiece
with that provided with the autofocus unit.
The MM6-OCC10X incorporates a focusing plate. Take
CAUTION
care not to leave dirt or stains on the focusing plate when
replacing the crossframe.
Also, when detaching and attaching the cross-frame unit,
stand the eyepiece so that the focusing plate does not drop.
Replacing the cross-frame
d
1
Remove the cross-frame c built into the MM6-OC10X or MM6-
OCC10X eyepiece by rotating the cross-frame.
2
Attach the cross-frame provided with the autofocus unit to the eyepiece.
c
} With the MM6-OCC10X, drop the focusing plate d into the cross-
frame so that the focusing plate's indication surface faces down,
and then screw the cross-frame onto the eyepiece.
STM7
67