Opening The Door Without Power - Thermo Scientific VARIOKLAV 65 TC Operating Manual

Steam sterilizer
Table of Contents

Advertisement

17 Opening the door without power

DANGER
Only use the emergency unlocking aid to open the sterilizer
door when the sterilizing chamber is not under pressure.
♦ Check to ensure that the sterilizing chamber is not under pressure.
Slightly jolt the door. If it is possible to move the door back and forth a bit, the
sterilizing chamber is not under pressure.
♦ Remove the cover for the emergency unlocking aid insertion opening.
♦ Insert the emergency unlocking aid in the emergency unlocking
aid insertion opening (see Figure 28).
Figure 28: Emergency unlocking aid insertion opening
♦ Fasten the hook of the emergency unlocking aid to the eye of the lock
barrel (not visible from outside).
♦ Pull the emergency unlocking aid.
The door opens.
♦ Remove the emergency unlocking aid and attach the insertion cover.
When the sterilizer is reconnected to mains, the sterilization phase display may
show an error message.
♦ Proceed as described in Chapter 14, Errors.
DANGER
If a program is interrupted, the load must be re-sterilized for
safety reasons.
Thermo Scientific
17 Opening the door without power
Eye
Lock barrel
82
50112712_082007_GB