Figure 13: Steel wire baskets
loaded with primary packs
♦ Place the items to be sterilized in the steel wire baskets.

8.3.3 Sterilization containers

INFORMATION
Type testing of the VARIOKLAV Type 65 TC benchtop sterilizer
was performed using AESCULAP and WAGNER sterilization
containers. We recommend using either AESCULAP or
WAGNER sterilization containers to ensure that the same
sterilization conditions prevail as during type testing.
INFORMATION
For information on using the respective sterilization containers
(filter changes etc.) and on storage times please consult the
manufacturer's instructions (AESCULAP, WAGNER).
The sterilization containers meet the requirements of the EN 868-8 standard as per
August 1999 and are suitable for accommodating items during vacuum steam
sterilization. The seal between the lid and the bottom has microbial barrier
properties, ensuring that sterility is maintained for as long as the items are
transported and stored under appropriate hospital conditions.
Perforated aluminium
floor with filter
Figure 16: Principal design of a sterilization container
Thermo Scientific
8 Preparing for sterilization
Paper
(back)
Foil
(front)
Figure 14: Structure
of primary packs
Perforated stainless
steel lid with filter
Quick lock
53
Free air and
steam circulation
Figure 15: Primary packs
in steel wire baskets
50112712_082007_GB