Thermo Scientific VARIOKLAV 65 TC Operating Manual page 30

Steam sterilizer
Table of Contents

Advertisement

Feed water tank
5
with glass cover and
handle
Control panel
6
7
Door seal
Anchors for locking
8
levers
9
Sterilizing chamber
10 Rack system
11 Feed water tank drain
12 Condensate gutter
13 Insert tray
14 I-M-S printer display
15 Current values display
key ▲
16
key ▼
17
18 Cycle display
19 Program keys
20 Green START/STOP key : Turns the sterilizer on from standby and starts the
21 Door indicator
Thermo Scientific
6 Equipment description
: The feed water tank contains the feed water
(deionized water) used to generate the sterile
steam. The glass cover facilitates visual
inspection at any time (feed water content) and
protects the feed water from contamination.
The spent-steam condenser subsequently
condensates the steam, which can then be
disposed.
: Keys and switch keys for programming; displays.
: Seals the sterilizer door.
: The locking levers keep the closed door tightly
locked to the housing anchors.
: Chamber in which items are sterilized.
: A shelf (optional accessory) can be inserted
centrically to split the sterilizing chamber.
: Connector for draining the feed water tank. A
piece of hose with a suitable coupling is included.
: Catches condensate that has formed on the door
or in the chamber.
: For placing sterilization containers in the
sterilizing chamber.
: Shows the current values of the sensors of the
I-M-S printer in the sterilizing chamber:
Chamber temperature (in °C)
Absolute pressure inside the sterilizing chamber
(in kPa).
The admission temperature of the cooling water
(vacuum pump) is also displayed below the
chamber temperature.
: Shows the current values of the control sensors
in the sterilizing chamber:
Chamber temperature (in °C)
Absolute pressure inside the sterilizing chamber
(in kPa)
Absolute pressure inside the double wall (in kPa)
Remaining sterilizing or drying time (in minutes)
: Adjustment key for setting longer drying times
and for selecting programs.
: Adjustment key for selecting a program.
: Shows set temperature, sterilizing time, and the
name of the selected program.
: Select the sterilizing program.
sterilization process.
: Indicates the condition of the door:
yellow
door closed, locked.
green
door open, unlocked.
While the door is being locked or unlocked by the
locking levers, both indicator lights are off.
30
50112712_082007_GB

Advertisement

Table of Contents
loading