Preparing For Sterilization - Thermo Scientific VARIOKLAV 65 TC Operating Manual

Steam sterilizer
Table of Contents

Advertisement

8 Preparing for sterilization

LIFE HAZARD
No matter what the problem, never use force when opening or
closing the swing door of the sterilizer. LIFE HAZARD!
The system might be damaged!
DANGER
Only run the sterilizer for correct use.
DANGER
Take note of all warning signs attached to the sterilizer.
DANGER
If problems occur that might be harmful to persons or
property, the unit must be immediately disconnected from the
power supply!
Disconnect the feed water supply where appropriate.
DANGER
Should any anomalies and / or problems occur while the
sterilizer is ready for use or while running a sterilization cycle:
Press the START/STOP key to terminate any ongoing
programs, if necessary.
CAUTION
Pursuant to DIN 58946-6, the sterilizer operator is obliged to
keep a device log to document important parameters and
events.
CAUTION
Take care to ensure that the door seal (7) is: free from dirt, not
sticking to the door (1), and not damaged.
The door seal (7) must sit snugly in the groove.
Thermo Scientific
8 Preparing for sterilization
35
50112712_082007_GB