CAUTION
A mains circuit breaker with an ELCB switch and the CEE
socket must be installed in the immediate vicinity of the unit.
CAUTION
Be careful when connecting media supply lines.
Danger of water-related damage!
CAUTION
Do not deploy the unit in the direct vicinity of a fire alarm.
Exiting steam could set off the fire alarm.
CAUTION
Take care to ensure that the door seal (7) is:
free from dirt, not sticking to the door (1), and not damaged.
The door seal (7) must sit snugly in the groove.
CAUTION
The equipment must not be operated in hazardous locations!
CAUTION
Before sterilization, check to see whether there is enough feed
water in the feed water tank (5).
Only use deionized water when topping up the feed water.
Do not use tap water!
CAUTION
Only use deionized or distilled water when topping up feed
water. Do not use tap water!
Feed water quality pursuant to DIN EN 285.
CAUTION
For error-free operation and to avoid corrosion damage,
carefully follow the maintenance manual.
Thermo Scientific
2 Safety considerations
12
50112712_082007_GB