Preparations For Transport (Frost Protection) - Thermo Scientific VARIOKLAV 65 TC Operating Manual

Steam sterilizer
Table of Contents

Advertisement

15 Preparations for transport
(frost protection)
♦ Start the Empty Jacket program (P
The water from the double wall is now drained into the disposal system. This
process may take some time, depending on the operating state.
♦ Provide a collecting vessel (container,
bucket). Place the free hose end into the
collecting vessel.
♦ Press the coupling into the coupling socket
(1).
When the feed water tank has been
drained via the silicone hose:
♦ Press against the outer edge of the
coupling socket (1) with force: The
coupling disengages from the socket.
♦ Clean the removable drip ledge (2).
♦ Remove the residual water from the tank.
♦ Clean the sieve of the feed water tank under running water.
♦ Clean the entire sterilizing chamber as well as the feed water tank
thoroughly with a moist soft cloth.
Draining the vacuum pump
♦ Place the vessel below the screw cap (see picture).
Thermo Scientific
15 Preparations for transport
(frost protection)
program key).
3
78
2
1
Screw cap
50112712_082007_GB