DANGER
Should the sterilizer remain unused for more than three hours
and cool down, a renewed sterilization check is recommended.
Perform the Vacuum & Empty and Bowie & Dick test programs.
DANGER
If the safety valve does not react to a higher pressure than the
set nominal pressure
Do not use it again until a new and fully functional safety valve
is installed.
DANGER
For more detailed information concerning maintenance consult
the maintenance instructions in the ring binder supplied with
the sterilizer.
DANGER
Repairs and interventions may only be carried out by our
authorized after-sales service.
DANGER
Any repairs and interventions concerning the sterilizer must be
documented in the device log.
DANGER
If an error message is displayed during the sterilization
process, the load must be considered non-sterile.
DANGER
Only use the emergency unlocking aid to open the sterilizer
door when the sterilizing chamber is not under pressure.
DANGER
The supply voltage and supply frequency must be the same as
given on the type plate.
Only connect the sterilizer to the specified socket.
DANGER
When a fault occurs and for maintenance purposes,
disconnect the steam sterilizer from the mains supply.
Pull the mains plug or neutralize the fuse.
)