DANGER
If the indicator strip of the test object with indicator system
(helix test) does not reveal four uniformly blackened bars,
optimal sterilizing conditions are not ensured. In this case,
please contact a service agency without delay and stop using
the sterilizer.
DANGER
If the indicator test sheet of the STERICLIN Bowie & Dick
single-use test pack does not change its colour uniformly and
thoroughly, optimal sterilizing conditions are not ensured.
In this case, please contact a service agency without delay and
stop using the sterilizer.
DANGER
If the Bowie & Dick test program is interrupted for some
reasons, a new test must be performed for safety reasons.
Make sure that the test frame and the test pack are in non-
heated condition and replace the Bowie & Dick indicator test
sheet.
There must be no water droplets inside the rolled-up helix test
tube (i.e. the test object with indicator system).
DANGER
Make sure that no cloth ends dangle from the sterilization
container when the items are wrapped in order to avoid non-
sterile batches.
DANGER
Each time a sterilization container is to be used, the paper filter
in the lid and (where appropriate) the bottom is to be checked.
This will guarantee safe storage conditions. Replace the filter if
necessary.
DANGER
Check the sterilized items for apparent damage. Damaged and
/ or moist packaged items must be considered non-sterile.