Thermo Scientific VARIOKLAV 65 TC Operating Manual page 9

Steam sterilizer
Table of Contents

Advertisement

DANGER
Mobile phones (cell phones) may interfere with the execution
of programs and therefore must be turned off at the installation
site or near the control unit of the sterilizer.
DANGER
Repairs and modifications are restricted to specially
authorized.
DANGER
Take note of all warning signs attached to the sterilizer.
DANGER
Perform repeated checks to assure the quality of sterilization.
The purpose of these checks is to protect both patients and
personnel. Observe the maintenance instructions in the ring
binder supplied with the sterilizer.
DANGER
Pursuant to DIN 58946-6, the sterilizer operator is obliged to
keep a device log to document important parameters and
events.
DANGER
Perform batch documentation and sterilizer release
documentation pursuant to DIN 58946-6 (June 2000).
DANGER
Never use test media (Bowie & Dick simulation test pack, batch
control indicator for helix test) whose shelf life has expired.
DANGER
Perform the Vacuum & Empty and Bowie & Dick test programs
once daily.
This allows you to check sterilizing conditions and sterilization
quality on a daily basis.
DANGER
To ensure uniform testing conditions, it is recommended to
perform the Bowie & Dick test when the sterilizer is hot. To
insure this, execute the Vacuum & Empty program first.
Thermo Scientific
2 Safety considerations
9
50112712_082007_GB