The locking levers of the door latch themselves for transport.
♦ Pull the mains plug.
INFORMATION
For information on the storage conditions, please refer to
Technical Specifications.
INFORMATION
It might be necessary to execute recommissioning according
to valid regulations (German Federal Armed Forces).

15.1 Recommissioning

Every intervention at the sterilizer and / or its supplies, such as changes, rebuilding
and retrofitting that may result in influencing the performance of the sterilizer,
require recommissioning.
When recommissioning, the commissioning tests are checked and, if necessary,
repeated.
Following tests must be executed at commissioning:
Concordance with specifications
Calibration of the units concerning control, display and documentation
Checking measuring, controlling and regulating features
Checking operational safety
Checking vacuum-tightness
Checking consumables
Visual inspection
Thermo Scientific
15 Preparations for transport
(frost protection)