B. Installing the Emergency Switch
● Install the emergency switch stop switch in the power cable line.
● Install this switch so that it is easy to reach in the emergency case but can't be pushed by
mistake.
● The switch shall be a type of mistake-proof.
● The switch is not supplied.
● The switch shall be installed at a height of 1.2 to 1.5 meters(47 to 60").
Emergency shutdown
1. The cable sizes: N, L and PE≥ 12 AWG(3 x 4 mm
2. The cable to emergency switch shall be the same size as the power cable itself.
3. Install the socket connector terminal for the 2nd protective ground wire.
Pax-i3D Smart(PHT-30LFO) Installation Manual
To PaX-i3D Smart
Plug
L
N
PE
Second Protective ground wire
L
N
PE
).
2
Centra distribution panel
w/ 25A circuit breaker
Appendix
147