Appendix
A. Installing the Warning Lamp and Door Interlock Switch
Requirements:
1. The warning control system shall be connected to the ERB (earth reference bar) of the room
that it is associated with.
2. The switching arrangements, location, height, and number of illuminated warning signs shall be
agreed with the local radiation protection advisor (RPA).(customer)
3. A fluorescent lamp shall not be used in the 'X-rays on' sign.
4. The customer shall be responsible for the proper installations for the warning control system,
including the lamp and door interlock switch, based on the MEIGaN guideline.
MEIGaN: Medical Electrical Installation Guidance Notes
5. Pre-installation planning is crucial to a successful installation for these devices.
6. For the further details, refer to the accompanying volume: Specification for Electrical Installation
Block diagram:
PaX-i3D Smart
144
Pax-i3D Smart(PHT-30LFO) Installation Manual
Need help?
Do you have a question about the PaX-i3D Smart and is the answer not in the manual?
Questions and answers