6. Select 100 x 70 in the FOV drop-down list box.
7. Subtract -7.1(mm) from FOV 10 x 85 Top position value and type the calculation value in the
Colm-T box. Ensure the other values(for Bottom/Left/Right) are the same as FOV 100 x 85
values and do not change these values.
8. Click Initial and wait until the system is initialized.
9. Click Capture when it is enabled.
10. When Align Sequence window appears, press the X-ray exposure switch
11. When Raw Viewer window appears, release the X-ray exposure switch.
12. Select 0009.raw in the raw file list.
13. Type 75 in the T box
14. Select S in the M box.
15. Make sure that the result meets the CBCT Top collimator alignment standard.
16. When the Collimator Alignment is completed , click Close to exit the Collimator Align program.
Pax-i3D Smart(PHT-30LFO) Installation Manual
If the Adult alignment Top value is -2.840, type -9.940 in the Colm-T
box.
(-2.840 - 7.100= -9.940)
Stay outside of the X-ray shielding room during the exposure.
9 Acquiring the test image