ZAA2600 - Manual
10.5 Status messages
10.5.1 Overload
Figure 1:Main screen - Film Applicator
Figure 2: :Main screen - Film Applicator
10.5.2 End switch
This message can only appear in film testing (scrubbing/crocking):
23. September 2021
"Motor has stopped because of overload. "This message
is only an information, you can tap either on "OK" or move
forward with
message be displayed repeatedly during application, the
applicator/scrubbing/crocking resistance is probably too
high.
A possible solution could be to increase the acceleration
mode. Select "Setup" and "Motor Settings", then
"Acceleration mode" and tap „Elevated" in the menu.
„Motor has stopped because end switch was reached".
The preset stroke length exceeds the position of the end
switch. It is recommended to always move the end stop
(2) completely to the front and the start stop (1)
completely to the back part of the device.
Solution: Move the end stop (2) completely to the front of
the device and press „OK" or „Scrubbing"
afterwards.
© Copyright 2021, PROCEQ SA
or
(respectively). Should this
68