Error Messages; General - Proceq PAROTESTER 2 Operating Instructions Manual

Paper and film tester
Table of Contents

Advertisement

«L-value too low»
Fig. 60
«L-value too high»
Fig. 61
«Max. no. of impacts»
Fig. 62
© 2012 Proceq SA
The value is below the hard-
ness range L<120. Such values
are neither displayed, recorded
or stored (Fig.60).
Stop measurement or the
measurement series with the
«END» key.
The value is above the hard-
ness range L >950. Such val-
ues are neither displayed nor
recorded or stored. Stop mea-
surement or the measurement
series with the «END» key.
(Fig. 61).
This message means that the
maximum number of measured
values per roll is reached
( n = 200) (Fig.62) No other
measured values are displayed,
recorded or stored.. Stop mea-
surement or the measurement
series with the «END» key.
«Memory full»
Fig.63
Warning
If measurement is continued, further measurements over-
write the saved measured values at the beginning of the
memory. The old measured data are then deleted from
the memory.
10.

Error messages

The impact energy of the impact device is always
monitored by the electronic system itself. The encoder
geometry and correct fit of the support ring are also
checked.
Error messages
When the display device is
switched on, the message
«Memory full» (Fig.63) can
appear on the screen when
the device is ready for mea-
surement. The measured data
should be uploaded to the PC
at the latest when this message
appears! Refer to section 7.
23