Disassembly Zaa2600.Sw - Proceq ZAA2600 Instruction Manual

Automatic film applicator and testing system
Table of Contents

Advertisement

ZAA2600 - Manual
Remove both holding device (8) and the clamping device (5) by
unscrewing their fixing screws.
Remove the clamping device (5), the inserted plate, the back heating
plate holder (21) and the front heating plate holder (22). (ZAA2600.HA
only)
Remove the isolation plate (20). (ZAA2600.HA only)
Remove the glass plate.
Mount both side panels (17) again, the one with slit on the right side,
the other one on the left side.
Refasten the previously removed screws of the start- (1) and end stop
(2).
Slide both stops (1 & 2) to the middle of the device.
Screw in the four black rubber feet. (ZAA2600.HA only)
Only transport the device in its original packaging.
Disassembled device before and after transport:

5.2.2 Disassembly ZAA2600.SW

Always switch off the device and unplug the power cable before dissembling.
To prevent transport damage, the device must be disassembled as follows:
Fold up the tool holder (4) and loosen the screws of the sample frames (6) with a quarter-
turn.
Remove the sample frames (6) and afterwards the sample tray (3).
Replace the sample frames (6) and fix them again, fold down the tool holder (4).
Move the tool holder (4) to the centre of the device.
Move the start stop (1) and the end stop (2) to the centre of the device as well and fasten
them with their screws.
Shipping condition:
23. September 2021
© Copyright 2021, PROCEQ SA
(16)
31