Should the applicator rod rotate during the application, use the optional ACC1348 fixing
unit for profile rods and wire-bar applicators (see chapter 6.2.2 on page 37)

6.2.1 Using the optional collecting device ACC1283

Using the collecting device, the application steps required when working with wire-bar
applicators and profile rods will be reduced significantly.
When using the collecting device, the applicator rod is lifted automatically off the substrate
after application and the drawing unit (20) moves back to the starting position automatically.
The applicator rod can then be easily removed and cleaned.
Use the collecting device as follows:
Make sure that the printing blanket side of the glass plate is facing up.
Ensure that the printing blanket is covered with the polyester mat.
Set the „Moving Mode" to „Reverse", by tapping on „Menu" and „Setup" on the touchscreen.
The „Reverse" mode is indicated by a bent arrow
Lift the weight (7) and hinge it into the top position.
Set the starting position as follows:
Place your substrate (e.g. test chart / film) at the desired position on the glass plate and fix
it with the clamping device (5), by pulling on the clamping grip (3).
Move the drawing unit (20) to the start stop, by pressing the „Manual back" button
Zero the device by tapping the „Zero"
Place the applicator rod into the holding device (8).
Adjust the collecting device as follows:
Place the collecting device into the end stop (2) by sliding the mounting pins of the device
into the holes of the end stop (2). Lock the collecting device in place by applying a light
pressure to it. Proper locking is indicated by a click.