15 Maintenance and Cleaning

15.1 Maintenance and cleaning work that can be carried out by the user

Only the following maintenance and cleaning work shall be carried out by the user:
Outer cleaning of the apparatus (see section 15.2)
Periodical lubrication of the guide shafts (see section 15.3)
Replacing of the fuse (see section 15.4)
Inspection (see section 15.5
All other maintenance and repair work shall only be carried out by Proceq SA or an
authorized Proceq Partner otherwise all guarantee and liability claims are void.

15.2 Cleaning

In order to ensure a perfect function, the film applicator should be kept as clean as possible.
Remove eventual stains of coating materials or adhesives immediately before they can dry.
Later they can often only be removed with difficulty.
For cleaning of the apparatus use commercially available cleaning agents such as cleaning
agent for glass, benzine, acetone. Do not use strong acids or alkaline liquids!
For cleaning of the printing blanket use only cleaning agents which do not affect the printing
blanket. Unsuitable for this purpose are solvents containing ketone such as acetone, or nitro-
cellulose diluents.
Before cleaning the ZAA 2600, always switch off the apparatus and unplug it. Never
immerse the apparatus in water or other liquids: Danger of short circuit.
While cleaning, take care that no cleaning liquid penetrates the interior of the
apparatus. The function of electrical or mechanical components could be impaired.
The edge of the printing blanket of the glass plate must not get wet.
If the device has to be disinfected, do not use disinfectants that contain sodium
hydroxide.
Never touch the hot plate, always wait until the heating plate has cooled.

15.3 Lubrication of the guide rods

From time to time it can be necessary to lubricate the guide rods with some sewing machine
oil or something similar.
Before lubricating the guide rods of the ZAA 2600, always switch off the apparatus and
unplug it.
No oil shall be sprayed into the apparatus. The function of electrical or mechanical
components could be impaired.