8.4.3 Mounting the friction finger set C

1) Check the cellular rubber for
damages.
3) Clamp the rubbing cloth smoothly. 4) Slide on the weight and screw it in place.
If your friction finger set does not look like picture 3, meaning one corner of the
ACC1353 rubbing cloth (circled in red) does not face the holder screw (circled in
green), you need to redo step 2.
The test load of 9 N will only be reached if the knurled nut is mounted.
To avoid ejection of weights (accidentally falling down) they have to be screwed to the
holder.

8.5 Carry out Crockmeter tests

Never touch any moving parts during operation.
The side panels must always be mounted during operation.
Do not lean over the unit during operation.
Risk of injury: Never hold the sample by hand during operation. It has to be fixed with
an appropriate means, e.g. double sided adhesive tape.
Crockmeter tests are carried out in machine mode „Crock/Scrub". Press menu on the
touchscreen -> Setup -> Machine mode, choose „Crock/Scrub", the sign
in the display.
23. September 2021
2) Place the rubbing cloth on the cellular
rubber with one corner (circled in red)
facing the holder screw (circled in green).