ZAA2600 - Manual
No friction sets must be inserted into the tool holder (4) for shipping.
5.2.3 Disassembly of the ZAA2600.C
Always switch off the device and unplug the power cable before dissembling.
To prevent transport damage, the device must be disassembled as follows:
Remove all friction finger sets from the tool holder bridge (4).
Remove the start stop (1) and the end stop (2).
Loosen the fixing screws of the side panels (16) and remove the side panels (15).
Remove the tool holder bridge (4) and the tool holder side parts (5).
Remount the side panels (15), the one with slit on the right side, the other one on the left
side.
Remount the start stop (1) and the end stop (2), move them to the centre of the device and
fasten them.
Remove the glass plate.
Shipping condition:
The glass plate (3) must be removed for shipping.
5.3 Startup
Connect the device to a power socket with the supplied power cable and turn on the main
switch (15).
Switch on the device by pressing the power button (10). The device will be ready after a
short start-up phase.
23. September 2021
© Copyright 2021, PROCEQ SA
32