7 Scrub test

7.1 Preparation

Before the apparatus is prepared for scrub tests, you should have decided:
according to which standard the scrub test should be carried out
which friction set you will be using
if an even or a curved sample will be inserted
Up to 4 identical or different friction sets can be used at the same time, depending on the
friction material, the sample surface, the speed and the test load.
The highest possible test load is depending on the friction material, the sample surface and
the speed.
The provided delivered glass plate is used for application as well as with the modification
sets.
Generally, the scrub tester can be used with the glass plate, though it is recommended to
use the scrub tester with the sample tray for scrub tests.
The following needs to be considered:
the thickness of the sample will be reduced due to the maximal height of the tool holder (4)
no liquids must be used.

7.2 Use of the friction sets

For an overview on the available friction sets and the content please see section 3.2.3.
The weight of the friction sets will only be reached if the knurled nut is mounted.
To avoid ejection of weights they have to be screwed to the holder.
Use of the ACC1468 friction sets for ISO 11998
Press the abrasive pad firmly onto the Velcro surface of the holder.
No additional weight is needed for tests according to ISO 11998.
Spare abrasive pads: ACC1483