Cycle Speed And Stroke Length Cannot Be Set; The Motor Is Overheated; Lubrication The Guide Rods - Proceq ZAA2600 Instruction Manual

Automatic film applicator and testing system
Table of Contents

Advertisement

ZAA2600 - Manual

10.5.3 Cycle speed and stroke length cannot be set

This message can only appear in film testing (scrubbing/crocking):

10.5.4 The motor is overheated

10.5.5 Lubrication the guide rods

23. September 2021
The fields for cycle speed and stroke length are marked
red. After tapping on the information symbol
message will appear that either the cycle counts or the
cycle speed are too high.
Solution: The motor cannot reach the preset number of
double strokes with the preset stroke length and speed.
Either the double stroke (cycle) speed or the stroke
length has to be reduced. Alternatively, the acceleration
mode or speed mode of the motor can be adjusted. For
this see section 10.4.
If the motor of the unit is overheated, this message
appears and the current action will be interrupted
automatically until the motor has cooled down. Confirm
the message with OK, the device switches off. Let the
device cool down before switching it on again. Arrange for
enough ventilation around the device. The unit must not
be operated outside of the allowed temperature range.
After a service performance of about 2'000 km a
message will appear at startup reminding you of to
lubricate the guide rods. Carry out the lubrication as
described in section 15.3.
To switch off the reminder until the next interval choose
"Do not show this message again" and close with OK.