Reset; Status Messages; Overload; End Switch - Proceq ZAA 2600.SW Instruction Manual

Automatic scrub tester
Table of Contents

Advertisement

7.5 Reset

7.6 Status messages

7.6.1 Overload

„Elevated" in the menu.

7.6.2 End switch

7.6.3 Cycle speed and stroke length cannot be set

With the „Reset" button the cycle
counter (double stroke counter) will be
reset. This cannot be undone.
If the „Reset" button has been tapped
this message appears which has to be
either confirmed or rejected.
"Motor has stopped because of
overload. "This message is only an
information, you can tap either on "OK"
or move forward with
the message be displayed repeatedly
during a procedure, the scrubbing
resistance is probably too high. A
possible solution could be to increase
the acceleration mode. Select "Setup"
and "Motor Settings", then
"Acceleration mode" and tap
„Motor has stopped because end
switch was reached". The preset stroke
length exceeds the position of the end
switch. It is recommended to always
move the end stop (2) completely to the
front and the start stop (1) completely
to the back part of the device.
Solution: Move the end stop (2)
completely to the front of the device
and press „OK" or „Scrubbing"
afterwards.