Noise Rejection - SIGLENT SDS1000X-E Series User Manual

Digital oscilloscope
Hide thumbs Also See for SDS1000X-E Series:
Table of Contents

Advertisement

SDS1000X-E&SDS1000X-U User Manual

Noise Rejection

Noise Reject adds additional hysteresis to the trigger circuitry. By increasing the
trigger hysteresis band, you reduce the possibility of triggering on noise. However,
this also decreases the trigger sensitivity so that a slightly larger signal is required
to trigger the oscilloscope.
Press the Setup button on the front panel, and then press the Noise Reject softkey
continually to set the option to On or Off to turn on or off the noise rejection
function.
Figure 28 Turn off the Noise Reject
66
WWW.SIGLENT.COM