SDS1000X-E&SDS1000X-U User Manual
type. It contains 6 types of Vpp, Vrms, dBV, dBu, dBm, Arbitrary dB.
When the unit type is dB, it is necessary to set the reference level.
Press the Load soft key to switch the load between 50Ω, 75Ω, 600Ω,
Hi-Z. It also can turn the Universal Knob to set the load value.
5)
Press the Set Channel softkey to enter the SET CHANNEL menu.
Press the DUT Input softkey to select the corresponding input signal
channel. Press the DUT Output 1 soft key to select the corresponding
output signal channel, and support up to 3 channels as the signal
output.
Press the Channel Gain softkey continuously to select the gain type as
"Hold" or "Auto".
2.
Press the Display soft key to enter the DISPLAY menu. The display menu
displays the configuration related to bode plot interface.
1)
Press the Amplitude softkey to enter the AMPLITUDE menu.
Press the Scale softkey and turn the Universal Knob to set the
amplitude display scale.
Press the Ref Level softkey and turn the Universal knob to set the
amplitude display reference level, or press the Universal knob directly
to enter the value in the pop-up virtual keyboard.
Press the Mode softkey to set the display mode of the amplitude.
Vout/Vin: display the ratio of the DUT output to the DUT input. Press
the Axis Type softkey to set the vertical axis to logarithmic or linear.
Vout: display the DUT output voltage value directly. Press the unit
softkey to set the unit type to Vpp, Vrms, dBV, dBu, dBm, or Arbitrary
dB.
Press the Auto Set softkey to automatically adjust the vertical direction
parameters based on the Bode plot data.
2)
Press the Phase softkey to enter the PHASE menu.
Press the Scale (Ref Level) softkey and turn the Universal Knob to set
the phase display scale (phase display reference level), or press the
Universal Knob directly to enter the value in the pop-up virtual
keyboard.
Press the Unit softkey continuously to switch the phase unit to Deg or
Rad.
Press the Auto Set softkey to automatically adjust the vertical direction
parameters based on the Bode plot data.
3)
Press the Auto Set softkey to automatically adjust the vertical direction
parameters based on the Bode plot data.
Figure 132 DISPLAY Menu
207
WWW.SIGLENT.COM