Vertical Position; Coupling - SIGLENT SDS1000X-E Series User Manual

Digital oscilloscope
Hide thumbs Also See for SDS1000X-E Series:
Table of Contents

Advertisement

SDS1000X-E&SDS1000X-U User Manual

Vertical Position

Turn the VERTICAL Position Knob to adjust the vertical position of the channel
waveform. Turn the knob clockwise to increase the vertical position and the
channel waveform moves up while counterclockwise to reduce the vertical position
and the waveform moves down. Push the knob to set the vertical position of the
channel waveform to zero.
During the adjustment, the vertical position information Volts Pos displays at the
bottom of the screen. The table below shows the range of vertical position
according to the volt scale.
Oscilloscope
SDS1000X-E
SDS1000X-U

Coupling

Set the coupling mode to filter out the undesired signals. For example, the signal
under test is a square waveform with DC offset.
When the coupling mode set to DC: the DC and AC components of the signal
under test can both pass the channel.
When the coupling mode set to AC: the DC components of the signal under test
are blocked.
When the coupling mode set to GND: the DC and AC components of the signal
under test are both blocked.
Press the CH1 button on the front panel; then press the Coupling softkey and turn
the Universal Knob to select the desired coupling mode. The default setup is DC.
The current coupling mode is displayed in the channel label at the right side of the
screen. You can also press the Coupling softkey continuously to switch the coupling
mode.
Volt Scale
500 μV/div ~ 100 mV/div
102 mV/div ~ 1 V/div
1.02 V/div ~ 10 V/div
1mV/div ~ 200 mV/div
206 mV/div ~ 10 V/div
30
Range of Vertical Position
±2V
±20 V
±200 V
±2 V
±100 V
WWW.SIGLENT.COM