3.
Use the probe to connect the input terminal of CH1 of the oscilloscope and the
"Compensation Signal Output Terminal" of the probe.
4.
Press the Auto Setup.
5.
Observe the waveform on the display. In normal condition, the display should
be a square waveform as shown in the figure below:
6.
Use the same method to test the other channels. If the square waveforms
actually shown do not match that in the figure above, please perform "Probe
Compensation" in the next section.
WARNING
To avoid electric shock during the use of probe, please make sure that
the insulated wire of the probe is in good condition and do not touch the
metallic part of the probe when the probe is connected to high voltage
source

Probe Compensation

When the probes are used for the first time, you should compensate the probes to
match the input channels of the oscilloscope. Non-compensated or poorly
Figure 7 Function Inspection
7
WWW.SIGLENT.COM