SIGLENT SDS1000CFL Series User Manual

SIGLENT SDS1000CFL Series User Manual

Digital storage oscilloscope
Hide thumbs Also See for SDS1000CFL Series:

Advertisement

Quick Links

User Manual
SDS1000CFL Series
Digital Storage Oscilloscope
V 1.1
2012 SIGLENT TECHNOLOGIES CO,.LTD

Advertisement

Table of Contents
loading

Summary of Contents for SIGLENT SDS1000CFL Series

  • Page 1 User Manual SDS1000CFL Series Digital Storage Oscilloscope V 1.1 2012 SIGLENT TECHNOLOGIES CO,.LTD...
  • Page 3 SIGLENT Declaration Copyright © by SIGLENT TECHNOLOGIES CO,.LTD. All rights reserved. Contents in this Manual are not allowed to copy, extract and translate before being allowed by SIGLENT. SDS1000CFL User Manual I...
  • Page 4: Brief Introduction

    SIGLENT Brief Introduction Model: SDS1000CFL Series: SDS1072CFL, SDS1102CFL, SDS1202CFL,SDS1302CFL, SDS1074CFL, SDS1104CFL, SDS1204CFL,SDS1304CFL Characteristic: ● The volume of the oscilloscope is cabinet and it is portable ● 7” Color TFT LCD display ● 2/4 channels, Bandwidth: 70MHz-300 MHz ● Single real-time sampling rate is:2GSa/s, Equivalent sampling rate is 50GSa/s.
  • Page 5 SIGLENT Standard Accessories: ● 1:1/10:1 probe (2/4 PCS) ● Power Cable that fits the standard of destination country ● Qualified Certification. ● Guaranty Card ● User Manual ● USB Cable SDS1000CFL User Manual III...
  • Page 6 SIGLENT General Safety Summary Carefully read the following safety precautions to avoid person injury and prevent damage to the instrument and any products connected to it. To avoid potential hazards, please use the instrument as specified. Only qualified technician should perform service procedures...
  • Page 7 SIGLENT Safety Terms and Symbols Terms used on the instrument. Terms may appear on the instrument: DANGER: Indicates an injury or hazard that may be immediately happen. WARNING: Indicates an injury or hazard that may be not immediately happen. CAUTION: Indicates that a potential damage to the instrument or other property might occur.
  • Page 8: Table Of Contents

    SIGLENT Content Brief Introduction..................II Chapter 1 Accidence................1 1.1 Accidence of Panel and Display Information...............2 1.2 Function checking ........................5 Probe ..........................7 1.3.1 Probe Safety......................7 1.3.2 Probe Attenuation Setting ..................7 1.3.3 Probe Compensation ....................8 Chapter 2 Functions Instruction and Operation........9 2.1 Menu and control button ....................10...
  • Page 9 3.1 Prompting messages:......................111 3.2 Troubleshooting........................ 113 Chapter 4 Service and Support............115 4.1 Maintain summary ......................115 4.2 Contact with SIGLENT....................116 Appendix A: Specifications..............117 Appendix B: Default setup ..............122 Appendix C: Daily Maintain and Cleaning ......... 124 SDS1000CFL User Manual VII...
  • Page 11: Chapter 1 Accidence

    SIGLENT Chapter 1 Accidence SDS1000CFL Series Digital Storage Oscilloscope is mini-type and portable bench type instruments, which could be used for measuring as the GND voltage. This Chapter shows you how to operate following tasks: ◆ Accidence of panel and Display information ◆...
  • Page 12: Accidence Of Panel And Display Information

    It is important for you to understand the DSO’s front panel before operating it. The following contents are the brief introduction for the front panel function, which is useful to be familiar with the operation of the SDS1000CFL Series Digital Storage Oscilloscope in short time.
  • Page 13 SIGLENT 1.1.2 Back and Side Connections The following images show back and side panel connection locations. Picture 1.1-2 SDS1000CFL Back and Side panel 1. Security Lock Receptacle 2. Pass/Fail Output 3. LAN Port 4. USB Host and USB Device Connector 5.
  • Page 14 SIGLENT 1. Trigger status Armed. The oscilloscope is acquiring pre-trigger data. All triggers are ignored in this state. Ready. All pre-trigger data has been acquired and the oscilloscope is ready to accept a trigger. Trig’d. The oscilloscope has seen a trigger and is acquiring the posttrigger data.
  • Page 15: Function Checking

    SIGLENT 1.2 Function checking When you check whether or not the oscilloscope could work smoothly, please operate as following: 1. Power On the oscilloscope. Press “DEFAULT SETUP” to show the result of the self check. The probe default attenuation is 1X.
  • Page 16 SIGLENT Picture 1.2-3 4. Press “CH1” two times to cancel the channel 1, Press“CH2”,”CH3” or ”CH4” to change screen into channel 2, channel 3 or channel 4, reset the channel 2 as step 2 and step 3. 6 SDS1000CFL User Manual...
  • Page 17: Probe

    SIGLENT 1.3 Probe 1.3.1 Probe Safety A guard around the probe body provides a finger barrier for protection from electric shock. Picture 1.3-1 Connect the probe to the oscilloscope and connect the ground terminal to ground before you take any measurements.
  • Page 18: Probe Compensation

    SIGLENT Note. When the attenuation switch is set to 1X, the probe limits the bandwidth of the oscilloscope to 6MHz (according to Probe spec). To use the full bandwidth of the oscilloscope, be sure to set the switch to 10X 1.3.3 Probe Compensation...
  • Page 19: Chapter 2 Functions Instruction And Operation

    SIGLENT Chapter 2 Functions Instruction and Operation To use your oscilloscope effectively, you need to learn about the following oscilloscope functions: ◆ Menu and control button ◆ Connector ◆ Auto Setup ◆ Default Setup ◆ Universal knob ◆ Vertical System ◆...
  • Page 20: Menu And Control Button

    SIGLENT 2.1 Menu and control button Showing as the following picture: Picture 2-1 ■ Channel buttons(CH1,CH2,CH3,CH4): Press a channel button to turn that channel ON or OFF and open the channel menu for that channel. You can use the channel menu to set up a channel. When the channel is ON. The channel button is lit.
  • Page 21 SIGLENT set the trigger level to be halfway between the minimum and maximum voltage level automatically. This is useful when you connect a signal to the EXT TRIG connector and set the trigger source to Ext or Ext/5. ■ FORCE: Use the FORCE button to complete the current waveform acquisition whether the oscilloscope detects a trigger or not.
  • Page 22: Connector

    SIGLENT 2.2 Connector Picture 2-2 ■ Channel Connector (CH1, CH2,CH3, CH4): Input connectors for waveforms display. ■ EXT TRIG: Input connector for an external trigger source. Use the Trigger Menu to select the “Ext” or “Ext/5” trigger source. ■ Probe Component: Voltage probe compensation output and ground. Use to electrically match the probe to the oscilloscope input circuit.
  • Page 23: Auto Setup

    SIGLENT 2.3 Auto setup The SDS1000CFL Series Digital Storage Oscilloscopes have a Auto Setup function that identifies the waveform types and automatically adjusts controls to produce a usable display of the input signal. Press the AUTO button, and then press the menu option button adjacent to the...
  • Page 24 SIGLENT Table 2-2 Auto set the function item Function Setting Acquire Mode Adjusted to Sampling Display Format Display Type Set to Dots for a video signal, set to Vectors for an FFT spectrum; otherwise, unchanged Vertical Coupling Adjusted to DC or AC according to the input...
  • Page 25: Default Setup

    SIGLENT 2.4 Default setup The oscilloscope is set up for normal operation when it is shipped from the factory. This is the default setup. To recall this setup, press the DEFAULT SETUP button. The options, buttons and controls that change settings when you press the DEFAULT SETUP button, refer to appendix B.
  • Page 26: Vertical System

    SIGLENT 2.6 Vertical System The vertical control could be used for displaying waveform, rectify scale and position Vertical Position knob Volt/div knob Picture 2.6-1 2.6.1 CH1, CH2, CH3, CH4 Channel Table 2-1 CH1,CH2,CH3,CH4 function menu 1: Option Setting Introduction DC passes both AC and DC components of the input signal.
  • Page 27 SIGLENT Table 2-4 CH1, CH2,CH3,CH4 function menu 2: Option Setting Instruction Turn on invert function. Invert Turn off invert function. Setup 1MΩ as input impedance 1MΩ Setup 50MΩas input impedance Input 50 MΩ (SDS1304CFL/SDS1302CFL/SDS1204CFL/ SDS1202CFL cantain this function) Press this button to enter the “Digital Filter Digital Filter menu”.(See table 2-5)
  • Page 28 SIGLENT Setting up CH1, CH2,CH3,CH4 Channels Each channel has its own separate Menu. The items are set up separately according to each channel. 1. Choosing Coupling Take the CH1 for example; the tested signal is a sine wave signal with DC deflection: ●...
  • Page 29 SIGLENT Set BW to 20MHz BW Limit Symbol Picture 2.6-3 3. Adjust Sensitivity Vertical scale adjusting has Coarse and Fine two modes, Vertical sensitivity range is 2mV/div~5V/div scale. Take the CH1 for example: ● Press“CH1”→“Volts/Div”→“Coarse”. It is the default setting of Volts/Div, and it makes the vertical scaling in a 1-2-5-step sequence from 2mv/div, 5mv/div, 10mv/div to 5v/div.
  • Page 30 SIGLENT 4.Setting Probe Attenuation In order to assort the attenuation coefficient, you need to response in the channel operation Menu. If the attenuation coefficient is 10:1, the input coefficient should be set to 10X, so that the mistake of the Volts/div information and measure testing should be forbidden.
  • Page 31 SIGLENT 6. Using the Digital Filter Press “CH1”→“Next Page page1/3”→ “Filter”, display the digital filter menu. Select “Filter Type”, then select “Upper Limit” or “Lower Limit” and turn the “Universal” knob to adjust them. ● Press “CH1”→“Next Page page1/3”→ “Filter” →“Off”. Turn off the Digital Filter function.
  • Page 32: Using Vertical "Position" Knob And "Volt/Div" Knob

    SIGLENT 2.6.2 Using Vertical “Position” knob and “Volt/div” knob ■ Vertical “POSITION” knob 1. Use the Vertical “POSITION” knobs to move the channel waveforms up or down on the screen. This button’s resolution is variety as per the vertical scale.
  • Page 33 SIGLENT Table 2-7 MATH function menu: Introduction Function Setting Math operate between signal source Operation A and B. SourceA CH1-CH4 Select the CHX as SourceA SourceB CH1-CH4 Select the CHX as SourceB Invert the Math waveform. Invert Turn off Invert function.
  • Page 34 SIGLENT 1. FFT Spectrum Analyzer The FFT process mathematically converts a time-domain signal into its frequency components. You can use the Math FFT mode to view the following types of signals: ● Analyze the Humorous wave in the Power cable.
  • Page 35 SIGLENT that you want to analyze in the center eight divisions of the screen. The oscilloscope calculates the FFT spectrum using the center 1024 points of the time-domain waveform. ● Turn the “Volts/div” knob to ensure that the entire waveform remains on the screen.
  • Page 36 SIGLENT 3. Select FFT window Windows reduce spectral leakage in the FFT spectrum. The FFT assumes that the YT waveform repeats forever. With an integral number of cycles, the YT waveform starts and ends at the me amplitude and there are no discontinuities in the signal shape A non-integral number of cycles in the YT waveform causes the signal start and end points to be at different amplitudes.
  • Page 37 SIGLENT 5. Measuring an FFT Spectrum Using Cursors You can take two measurements on FFT spectrums: magnitude (in dB) and frequency (in Hz). Magnitude is referenced to 0 dB, where 0 dB equals 1 VRMS. You can use the cursors to take measurements at any zoom factor. (Refer to cursor measure2.11.2)
  • Page 38 SIGLENT Picture 2.6-11 Measure FFT Frequency Press the CURSOR button. Press the “Cursor Mode” button to select “Manual”. Press the “Type” option button to select “Time”. Press the “Source” option button to select “FFT”. Press the “CurA” option button, turn the “Universal” button to move Cursor A to the highest position of the FFT waveform.
  • Page 39: Using Ref

    SIGLENT 2.6.4 Using Ref The reference control saves waveforms to a nonvolatile waveform memory. The reference function becomes available after a waveform has been saved. Table 2-12 REF function menu: Introduction Option Setting CH1,CH2 CH3,CH4 CH1 off Source Choose the waveform display to store.
  • Page 40 SIGLENT 4. Press the third option button to select “REFA” ,”REFB”,”REFC” or “REFD” as storage position. 5. Press the “Save” option button. 6. Press the bottom option button to select “REFA ON” ,”REFB ON”,”REFC ON” or “REFD ON” to recall the reference waveform.
  • Page 41: Horizontal System

    SIGLENT 2.7 Horizontal System As follow Picture, there are one button and two knobs in the HORIZONTAL area. Horizontal S/div knob position knob Picture 2.7-1 Table 2- 13 SDS1000CFL Horizontal system function menu: Option Setting Description Turn on this function that main timebase...
  • Page 42: Window Zone

    SIGLENT ■ Display scan mode: When the SEC/DIV control is set to100 ms/div or slower and the trigger mode is set to Auto, the oscilloscope enters the scan acquisition mode. In this mode, the waveform display updates from left to right. There is no trigger or horizontal position control of waveforms during scan mode.
  • Page 43: Trigger System

    SIGLENT 2.8 Trigger System The trigger determines when the oscilloscope starts to acquire data and display a waveform. When a trigger is set up properly, the oscilloscope converts unstable displays or blank screens into meaningful waveforms. There are three buttons and one knob in the Trigger area.
  • Page 44: Signal Source

    SIGLENT 2.8.1 Signal Source You can use the Trigger Source options to select the signal that the oscilloscope uses as a trigger. The source can be any signal connected to a channel BNC, to the EXT TRIG BNC or the AC power line (available only with Edge triggers).
  • Page 45 SIGLENT Use this mode to let the acquisition free-run in the absence of a valid trigger; This mode allows an Auto untriggered, scanning waveform at 100 ms/div or slower time base settings. Use this mode when you want to see only valid Mode triggered waveforms;...
  • Page 46 SIGLENT Operate Instruction: 1. Set up Type 1) Press the “TRIG MENU” button to display “Trigger” menu. 2) Press the “Type” option button to select “Edge”. 2. Set up Source According to input signal, press the “Source” option button to select “CH1”, “CH2”,”CH3”,”CH4”,“EXT”, “EXT/5”or “AC Line”.
  • Page 47 SIGLENT Pulse Trigger:Use Pulse Width triggering to trigger on aberrant pulses. Table 2-16 Pulse Trigger function Menu 1: Option Setting Explain Select the pulse to trigger Type Pulse pulse match trigger condition. Source Select input signal source. EXT/5 (Positive pulse width less...
  • Page 48 SIGLENT Table 2-17 Pulse Trigger function Menu 2: Option Setting Explain Select the pulse to trigger the pulse match Type Pulse the trigger condition. Select the type of triggering; Normal mode is Auto best for most Pulse Width trigger Mode Normal applications.
  • Page 49 SIGLENT Video Trigger:Trigger on fields or lines of standard video signals. Table 2-18 Functional Manu of Video Trigger 1: Option Setting Instruction When you select the video type, put the couple set to the AC, then you could Type Video trigger the NTSC,PAL and SECAM video signal.
  • Page 50 SIGLENT Picture 2.8-5 Operate Instruction: 1. Set up Type 1) Press the “TRIG MENU” button to display “Trigger” menu. 2) Press the “Type” option button to select “Video”. 2. Set up Polarity Press the “Polarity” option button to select “...
  • Page 51 SIGLENT Slope Trigger: Trigger on positive slope of negative slope according to setup time of the oscilloscope. Table 2-20 Slope trigger function menu 1 Option Setting Instruction Trigger on positive slope of negative Type Slope slope according to setup time of the oscilloscope.
  • Page 52 SIGLENT Table 2-21Slope trigger function menu 2 Option Setting Instruction Trigger on positive slope of negative slope Type Slope according to setup time of the oscilloscope. Vertical Select the trigger level that can be adjusted by “LEVEL” knob. You can adjust “LEVEL A”, “LEVEL B”...
  • Page 53 SIGLENT Press the “TRIG MENU” button to enter “Trigger menu”. Press the “Type” option button to select “Slope”. Press the “Source” option button to select “CH1”,”CH2”,”CH3” or “CH4”. Press the “When” option button to select “ ”, “ ”, “...
  • Page 54 SIGLENT Table 2-22 Set trigger mode to edge trigger function menu 1: Option Setting Instruction When use alternative trigger, the trigger signal Type Alternative comes from two vertical channels. In the mode, you can observe two irrelative signals at a time.
  • Page 55 SIGLENT Table 2-25 Set trigger mode to video trigger function menu1: Option Setting Instruction Type Alternative The trigger signal comes from two vertical channels when you use alternative trigger. In this mode, you can observe two irrelative signals at the same time.
  • Page 56 SIGLENT When Select slope trigger condition. Enter the second page of the alternative Next Page Page 1/2 trigger. Table 2-28 Set trigger mode to slope trigger function menu 2 Option Setting Instruction Time Turn the “Universal” knob to set the slope 〈...
  • Page 57: Coupling

    SIGLENT 2.8.3 Coupling Use the “Coupling” to make sure the signal that passes the trigger circuit. It is useful for us to gather a steady wave form. If you use the trigger coupling, you should press the “TRIGGER” button and then select “edge”, “Pulse”, “Video”...
  • Page 58: Trigger Holdoff

    SIGLENT 2.8.6 Trigger Holdoff You can use the Trigger Holdoff function to produce a stable display of complex waveforms. Holdoff is time between when the oscilloscope detects one trigger and when it is ready to detect another. The oscilloscope will not trigger during the holdoff time.
  • Page 59: Acquiring Signals System

    SIGLENT 2.9 Acquiring Signals system Showing as the follow picture, the “ACQUIRE” button for Acquiring Signals system is at the menu. Picture 2.9-1 Table 2-29 The Function manual of Acquiring Signals: Option Setting Introduction Use for sampling and accurately display most Sampling of the waveform.
  • Page 60 SIGLENT Disadvantage: This mode does not acquire rapid variations in the signal that may occur between samples. This can result in aliasing may cause narrow pulses to be missed. In these cases, you should use the Peak Detect mode to acquire data.
  • Page 61 SIGLENT Average: The oscilloscope acquires several waveforms, averages them, and ■ displays the resulting waveform. Advantage: You can use this mode to reduce random noise Picture 2.9-4 Average mode ■ Equivalent Time Sampling: The equivalent time sampling mode can achieve up to 20 ps of horizontal resolution (equivalent to 50GSa/s).
  • Page 62 SIGLENT 2). Continue to acquire data while waiting for the trigger condition to occur. 3). Detect the trigger condition. 4). Continue to acquire data until the waveform record is full. 5). Display the newly-acquired waveform. ■ Time Base: The oscilloscope digitizes waveforms by acquiring the value of an input signal at discrete points.
  • Page 63 SIGLENT “Sinx” is sine interpolation Picture 2.9-6 “X”is linear interpolation. Picture 2.9-7 Set up Sampling mode Press the “Mode” option button to select “Real Time” or “Equ Time”. Set up Sampling rate Adjust the sampling rate by pressing the “Sa Rate” option button and turning the Time/div front panel knob.
  • Page 64: Display System

    SIGLENT 2.10 Display System The display function could be expressed by the “DISPLAY” Button. Picture 2.10-1 Table 2-30 Display system function menu 1: Option Setting Introduction Vectors fills the space between adjacent sample points in the display. Dots: Vectors Type...
  • Page 65 SIGLENT Table 2-31 Display system function menu 2: Option Setting Introduction YT format displays the vertical voltage in relation to time (horizontal scale). XY Format format displays a dot each time a sample is acquired on channel 1 and channel 2, or channel 3 and channel 4.
  • Page 66 SIGLENT 2. Set up Persist Press “Persist” option button to select “Off”, “1 Sec”, “2 Sec”, “5Sec” or “Infinit ”. You can use this option to observe some especial waveforms. Picture 2.10-4 3. Set up Intensity 1) Press the “Intensity” option button and turn the “Universal” knob to adjust waveforms’...
  • Page 67: X-Y Format

    SIGLENT 2.10.1 X-Y Format Use the XY format to analyze phase differences, such as those represented by Lissajous patterns. The format plots the voltage on channel 1 against the voltage on channel 2, or plots the voltage on channel 3 against the voltage on channel 4, where channel 1(or channel 3) is the horizontal axis and channel 2(or channel 4) is the vertical axis.
  • Page 68: Measure System

    SIGLENT 2.11 Measure System The Oscilloscope displays the voltage in relation to time and test the wave form displayed. There are scale, Cursor and auto measure modes. 2.11.1 Scale Measurement This method allows you to make a quick, visual estimate. For example, you might look at waveform amplitude and determine that it is a little more than 100 mV.
  • Page 69 SIGLENT Manual Mode Table 2-33 Manual cursor function menu: Option Setting Instruction Cursor In this menu, set the manual cursor Manual Mode measure. cursor measure voltage Voltage Type parameters. Time Use cursor to measure time parameters. Source MATH/FFT Select input signal channel.
  • Page 70 SIGLENT “CH4”,“MATH”, “FFT” “REFA” ,“REFB”, “REFC”, “REFD” according to input signal channel. 5. Select “Cur A”, turn the “Universal” knob to adjust Cursor A. 6. Select “Cur B”, turn the “Universal” knob to adjust Cursor B. 7. The measurement values are displayed on the top of the left corner: If the measurement type is set to “Voltage”, the values are:...
  • Page 71 SIGLENT Track mode Table 2-34 Track mode function menu: Option Setting Instruction Cursor Mode Track In this mode, set track cursor measure. Set the input signal channel that the Cursor Cursor A A will measure. NONE Set the input signal channel that the Cursor Cursor B B will measure.
  • Page 72 SIGLENT B→T: The horizontal position of Cursor B (Time cursor centered around the midpoint of screen). B→V: The Vertical position of Cursor B (Voltage cursor centered around channel ground level). △T: Horizontal space between Cursor A and Cursor B (Time value between two cursors).
  • Page 73 SIGLENT Auto mode Table 2-35 Auto mode function menu: Option Setting Instruction Cursor Mode Auto Set to auto cursor measure mode. This mode will take effect with automatic measurements. The instruments will display cursors while measuring parameters automatically. These cursors demonstrate the physical meanings of these measurements.
  • Page 74: Auto Measurement

    SIGLENT 2.11.3 Auto Measurement Picture 2.11-5 When you take automatic measurements, the oscilloscope does all the calculating for you. Because the measurements use the waveform record points, they are more accurate than the graticule or cursor measurements. Press the ‘MEASURE’ for the Automatic Test.
  • Page 75 SIGLENT Table 2-37 Auto measure function 2-Voltage measure menu: Option Setting Instruction Select input signal source for Source CH1, CH2,CH3,CH4 Voltage measure. Vmax, Vmin, Vpp, Vamp, Press the “Type” button or Vtop, Vbase, Cycle Mean, turn the “Universal” knob to...
  • Page 76 SIGLENT Table 2-40 All Measurement function menu: Option Setting Instruction Source Select input signal channel. Turn on the all measurement function to measure voltage parameters. Voltage Turn of the all measurement function to measure voltage parameters. Turn on the all measurement function to measure Time parameters.
  • Page 77 SIGLENT Fall Time Measures the time between 90% and 10% of Fall Time the first falling edge of the waveform. The duration of a burst. Measured over the entire waveform. BWid + Width Measures the time between the first rising...
  • Page 78 SIGLENT If you want to measure voltage parameters, please follow next steps: 1). Press the “MEASURE” button to enter the “Auto measurement” menu. 2). Press the top first option button to enter the “second measurement menu”. 3). Select measure type. If you press the “Voltage” option button, “Voltage measurement”...
  • Page 79 SIGLENT If you want to measure time parameters using all measure function, please follow next steps: 1). Press the “MEASURE” button to enter the “Auto Measure menu”. 2).Press the top option button to enter the second page of “Auto Measure menu”.
  • Page 80: Storage System

    SIGLENT 2.12 Storage System Showing as following picture, The SAVE/RECALL is the Storage System Function Button. Picture 2.12-1 You can quickly save and recall up to 20 oscilloscope panel settings, 20 groups’ waveforms in internal memory. There is a USB Host interface in the front panel of the oscilloscope and you can save setup data, waveform data, waveform interface image, CSV file to a USB flash drive furthest at a time.
  • Page 81 SIGLENT While File shows option buttons for New File, Delete File, and Load. Picture 2.12-3 Recalling Files The Load button is used to recall your setup files. Once you’ve navigated to the desired file and it’s highlighted in the main screen area, press the Load option button and the setup is recalled from the USB flash drive.
  • Page 82 SIGLENT Creating Folders and Files Creat new folders and files by pressing the New Folder or New File option button. The following screen is shown. Picture 2.12-5 ● The New Filemenu choices and behavior is the same as the New Folder menu.
  • Page 83 SIGLENT SAVE/RECALL SETUP ■ Save Setups to Device: The complete setup is stored in nonvolatile memory. When you recall the setup, the oscilloscope will be in the mode from which the setup was saved. The oscilloscope saves the current setup if you wait three seconds after the last change before you power off the oscilloscope.
  • Page 84 SIGLENT 3)Press the “Save to” option button to select “Device”. 4)Press the “Setup” option button to select “No.1”. 5)Press the “DISPLAY” button to enter the “Display” menu. 6)Press the “Type” option button to select “Dots”. 7)Press the “SAVE/RECALL” button to enter the “SAVE/RECALL” menu.
  • Page 85 SIGLENT Save setup to USB flash drive, follow these steps: For example: Save setup that set waveform display type to “Dots” to USB flash drive. Press the “SAVE/RECALL” button to select “Setups”. 1) Insert USB flash drive to USB host port of the oscilloscope and wait that the oscilloscope has initialized USB flash drive (about five second).
  • Page 86 SIGLENT ■ RECALL FACTORY You can use this option to recall the factory setup. Table 2-44 Factory function menu: Instruction Option Setting To view the Factory setup. Type Factory Update the config file. Update cfg Recall the Factory setup. Load Picture 2.12-8...
  • Page 87 SIGLENT Picture 2.12-9 Save waveforms in internal memory, follow next steps: 1). Input a sine signal to channel 1 and press the “Auto” button. 2). Press the “SAVE/RECALL” button to enter “SAVE/RECALL” display menu. 3). Press the “Type” option button to select “waveforms”.
  • Page 88 SIGLENT ■ Save waveforms to USB flash drive Table 2-46 Save waveforms to USB flash drive function menu: Introduction Option Setup Menu for the Storage/Recall waveforms. Type Waveforms Save waveforms to USB flash drive. Save to File Accomplish the storage.
  • Page 89 SIGLENT Recall waveforms form USB flash drive, follow next steps: 1). Press the “SAVE/RECALL” button. 2). Press the “Type” button to select “Waveforms”. 3). Insert USB flash drive to USB host port of the oscilloscope and wait that the oscilloscope has initialized USB flash drive (about five seconds).
  • Page 90 SIGLENT Picture 2.12-11 Save waveform image to USB flash drive, follow next steps: 1). Select the screen image that you want . 2). Press the “SAVE/RECALL” button to enter “SAVE/RECALL” menu. 3). Press the “Type” option button to select “Pictures”.
  • Page 91 SIGLENT Save/Recall CSV Table 2-48 CSV storage function menu: Option Setting Introduction Type Menu for the Storage CSV file to USB flash drive. Data Displayed Set to store displayed waveform data to CSV file. Depth Maximum Set to store maximum waveform data to CSV file.
  • Page 92: Utility System

    SIGLENT 2.13 Utility System Picture 2.13-1 Table 2-49 Utility System function menu 1: Option Setting Introduction System Displays summaries of the oscilloscope Status settings. Open the key-press voice. Sound Close the key-press voice. Turn on Frequency Counter Counter Turn off Frequency Counter.
  • Page 93 SIGLENT Table 2-50 Utility System function menu 2: Option Setting Introduction Do self cal Auto self emendation. Screen Test Run the screen detect program Do Self Keyboard Test Run the keyboard detect program Test LED Test Run the dot lighten detect program.
  • Page 94 SIGLENT Table 2-52 IP Setting function menu: Introduction Option Setting Set oscilloscope IPAddress,adopt LAN port IP Address to connect the oscilloscope. Subnet Set corresponding IP subnet mask. Mask IP Setting Gate Way Set corresponding IP gate way. Port Connect port, adopt default port.
  • Page 95 SIGLENT Picture 2.13-5 Table 2-54 Function Menu of the Date/Time : Option Setting Introduction Set the present displaying year, change Year 2000-2099 the year by adjusting the universal knob. Set the present displaying month and Month/Day day, change the them by adjusting the 00/00 universal knob.
  • Page 96: System Status

    SIGLENT 2.13.1 System status Press the “System Status” option butto of the Utility Menu to view the oscilloscope’s hardware and software configuration. Picture2.13-6 Table 2-55 System status instruction: Option Introduction Startup Times List the boot-strap times. Software version List the software version.
  • Page 97: Language

    SIGLENT 2.13.2 Language The oscilloscopes have twelve languages’ user menu to be selected. Press the “Utility” button →“language” to select language. Picture 2.13-7 SDS1000CFL User Manual 87...
  • Page 98: Print

    SIGLENT 2.13.3 Print The scope support PictBridge compatible printers. You can connect the back USB port of the oscilloscope to the USB port of the PictBridge compatible printer using the USB cable. Table 2-56 Print setup option function menu 1:...
  • Page 99 SIGLENT Picture 2.13-8 Table 2-57 Print setup option function menu 2: Introduction Option Setting Default, 2.5x3.25 in, L (3.5x5 in), 4x6 Displays Settings in, 2L (5x7 in), 8x10 in, 4L (7 x 10 in), available on your Image Size E, Card, Hagaki card, 6 x 8 cm, 7x10...
  • Page 100 SIGLENT Picture2.13-9 To print screen picture, follow these steps: 1. Connect the oscilloscope to a PictBridge compatible printer. 1) Insert one end of a USB cable into the USB Device port on the oscilloscope. 2) Insert the other end of the cable into the PictBridge port on a PictBridge compatible printer.
  • Page 101: Self Calibration

    SIGLENT 2.13.4 Self Calibration Self Calibration procedure can optimize the signal path at the most measurement precision..You can run this procedure at any time. If the operating temperature changes by more than 5° C or the units run more than thirty minutes,you should do the self calibration.
  • Page 102 SIGLENT Operation Steps: 1. Screen Test: Select “Screen Test” to enter the screen test interface. The clew words “Press ‘SINGLE’ Key to continue,Press ‘RUN/STOP’ Key to exit” is displayed, you could press the “Single” for test. Picture 2.13-11 2. Keyboard Test Select “keyboard Test”...
  • Page 103 SIGLENT Picture 2.13-12 3. LED test Select “LED Test” to enter the lighten interface, the on-screen lathy rectangle shapes represent the front panel keys; the lathy rectangle shapes represent the front panel knobs. The squares represent the knob presses for scale knobs. The clew words “Press ‘SINGLE’...
  • Page 104: Updating The System Software

    SIGLENT 2.13.6 Updating the System Software ■ Using USB flash drive update frimware The software of the oscilloscope can be updated directly via USB Flash drive, this process needs about two minutes. Follow next steps: 1. Insert USB Flash Drive with firmware procedure to USB Host interface on the front panel of the oscilloscope.
  • Page 105 SIGLENT Picture 2.13-14 Table 2-60 Pass/Fail function menu 2: Option Setting Instruction Pass Output when pass condition detected. Output Fail Output when fail condition detected. Stop test when output occur. Stop On Output Continue test when output occur. Mask Setting Enter the “Mask Setting menu”.
  • Page 106 SIGLENT Table 2-61 Mask Setting function menu 1: Option Setting Instruction Turn the “Universal” knob to set the range of X Mask horizontal clearance waveform. xdiv <0.04div-4.00div> Turn the “Universal” knob to set the range of Y Mask vertical clearance waveform.
  • Page 107 SIGLENT Picture 2.13-17 RUN Pass/Fail test, please follow next page: 1) Press UTILITY button to enter the “Utility menu”. 2) Press the “Next Page Page1/4” option button. 3) Press the “Next Page Page2/4” option button to enter the third page of “Utility menu”...
  • Page 108: Waveform Record

    SIGLENT Picture 2.13-18 2.13.8 Waveform Record Waveform record can record input waveform from every channel with a maximum record length of 2500 frames. This record behavior can also be activated by the pass/fail test output, which makes this function especially useful to capture abnormal signals in long term without keeping an eye watching it.
  • Page 109 SIGLENT Picture2.13-19 To record waveforms, follow these steps: 1. Press the UTILITY button to enter “Utility menu”. 2. Press the “Next Page” option button to enter the third page of “Utility menu”. 3. Press the “Record” button to enter “Waveform Record Menu”.
  • Page 110 SIGLENT Picture 2.13-20 Table 2-65 Waveform play back function menu 2: Option Setting Instruction Start Frame Set start frame. Curr_Frame Select current frame to be played. End Frame Set End frame. Return Press to return the waveform recorder main menu.
  • Page 111 SIGLENT 4. Press the “Interval” option button to select time interval between play back frames. 5. Press the “Next Page 1/2” to enter the second page of “Play Back function menu”. 6. Select the “Start Frame” option, turn the “Universal” knob to adjust the start frame of play back waveform.
  • Page 112: Recorder

    SIGLENT 2.13.9 Recorder The waveform recorder is a kind of seamless and no-gap real time recording of waveform, means oscilloscope can save and replay waveform every time it captured. It is similar to waveform recording instrument, and the biggest recording size of its internal memory is 6M Table 2-66 waveform recording function menu:...
  • Page 113 SIGLENT Picture 2.13-23 Table 2-68 recorder’s setting menu Option Setting Description Full Screen Waveform of Full screen recording and channel recalling Waveform of recording on split screen Viewer Split channel recalling, displayed in upper half-screen, CH2 is displayed in lower half-screen.
  • Page 114: Remote Control

    SIGLENT Recorder Operation step: 1. Press the UTILITY button to enter “Utility menu”. 2. Press the “Next Page” option button to enter the four page of “Utility menu. 3. Press the “Recorder” button to enter the Recorder manual. 4. Press the “Option” button to select the parameter which you want by yourself 5.
  • Page 115 SIGLENT Use PC software offered by SIGLENT or other manufacturers User can send command to remotely control the oscilloscope through PC software EsayScope which offered by SIGLENT. Besides, “Measurement & Automation Explorer” which provided by NI (National Instruments Corporation) also can be used to control the oscilloscope.
  • Page 116 SIGLENT 106 SDS1000CFL User Manual...
  • Page 117 SIGLENT SDS1000CFL User Manual 107...
  • Page 118 SIGLENT 108 SDS1000CFL User Manual...
  • Page 119 SIGLENT Research Device Open EsayScope software; click “add device” to research, there will pop-up the following dialogue box, then click Add. Check Instrument Resource The instrument information which has been researched is as the picture below. It shows the instrument number and the USB interface information.
  • Page 120: Online Help Function

    SIGLENT 2.14 Online Help Function The oscilloscope has an online help function that supplies multi-language help information, and you can recall them to help you operate the oscilloscope when you need. Press the “HELP” button to enter the help status, then press every button to recall the corresponding help information.
  • Page 121: Chapter 3 Prompting Messages And Troubleshooting

    SIGLENT Chapter 3 Prompting Messages and Troubleshooting 3.1 Prompting messages: ■ Trig level at limit! : Mention you that the trigger Level is at a limit when you turn the Trig level knob. ■ Horizon position at limit! : Mention you that the horizontal position is at a limit when you turn the horizon position knob.
  • Page 122 SIGLENT internal of the oscilloscope or USB flash successful. ■ Please set Back USB to printer! : Press the “S/div” knob will appear this information on the screen when the “Print Key” option is set to “Print Picture” and the “Back USB” option is set to “Computer”.
  • Page 123: Troubleshooting

    SIGLENT 3.2 Troubleshooting 1. After the oscilloscope is powered on, the screen remains dark , please do as following steps: ⑴ Check the power cable’s connection. ⑵ Ensure the power switch is turned on. ⑶ After the inspections above,restart the oscilloscope.
  • Page 124 SIGLENT automatically. 6. After the Acquisition is set to Averages or Display Persistence time is set too long, the waveform refreshes slowly. It is normal in these settings 7. The signal is displayed as ladder like waveform (1)This phenomenon is normal. The time base maybe is too slow .you should turn the horizontal SCALE knob to increase horizontal resolution to improve the display.
  • Page 125: Chapter 4 Service And Support

    (3) years (accessories for a period of one year) from the date of shipment from an authorized Siglent’s distributor. If a product or CRT proves defective within the respective period, Siglent will provide repair or replacement as described in the complete warranty statement.
  • Page 126: Contact With Siglent

    SIGLENT 4.2 Contact with SIGLENT MTR Add:3//F, Bldg No.4, Antongda Industrial Zone, 3rd Liuxian Road ,Bao’an District, Shenzhen, 518101, P.R.China Tel:0086-755-3661 5186 E-mail:support@siglent.com http://www.siglent.com 116 SDS1000CFL User Manual...
  • Page 127: Appendix A: Specifications

    SIGLENT Appendix A: Specifications All specification applies to 10X probe and All the SDS1000CFL Series Digital Storage Oscilloscopes. To verify that the oscilloscope meets specifications, the oscilloscope must first meet the following conditions: ● The oscilloscope must have been operating continuously for thirty minutes within the specified operating temperature.
  • Page 128 SIGLENT offset range 102mV-5V: ±40V) Vertical Resolution 8 bit Channels 300MHz(SDS1304CFL, SDS1302CFL) Analog 200MHz(SDS1204CFL, SDS1202CFL) Bandwidth 100MHz(SDS1104CFL, SDS1102CFL) 70MHz(SDS1074CFL, SDS1072CFL) 300MHz(SDS1304CFL, SDS1302CFL) Single-shot 200MHz(SDS1204CFL, SDS1202CFL) Bandwidth 100MHz(SDS1104CFL, SDS1102CFL) 70MHz(SDS1074CFL, SDS1072CFL) DC -10% of rated BW: +/- 1dB BW Flatness at BNC...
  • Page 129 SIGLENT Sampling points: 1024 20MHz ± 40% (Note: BW limited below Bandwidth limited 20MHz when using probe in x1) Horizontal System Real Time Sampling Single Channel: 2GSa/s, Double Channels: Rate 1GSa/s (When timebase faster than 25ns/div) Equivalent Sampling The highest equivalent sampling rate of other...
  • Page 130 SIGLENT CH1 trigger type: Edge, Pulse, Video, Slope CH2 trigger type: Edge, Pulse, Video, Slope Alternative Trigger CH3 trigger type: Edge, Pulse, Video, Slope CH4 trigger type: Edge, Pulse, Video, Slope X-Y Mode X-pole Input / Y-Pole (CH1) / (CH2) or (CH3)/(CH4)
  • Page 131 SIGLENT Generic Specification Display System Color TFT 7.0in.(177.8mm)diagonal Display Mode Liquid Crystal Display Resolution 480 horizontal by 234 vertical pixels Display Color 64K color Display Contrast (Typical state) 150:1 Backlight Intensity(Typical state) 300nit Wave display range 8 x 18 div...
  • Page 132: Appendix B: Default Setup

    Appendix B: Default setup options 、 knobs or Menu or system Default setup buttons Coupling BW limit Volts/div Coarse CH1,CH2,CH3,CH4 Probe Invert Filter Volts/div 1.00V Operation SourceA SourceB Invert FFT operation: MATH Source Window Hanning FFT Zoom Scale dBVrms Display Split Window Main...
  • Page 133 Brightness Format Menu Display infinite Type Setups SAVE/RECALL Save To Device Setup No.1 REFA/REFB REFA Source REFA REFB Sound Counter UTILITY Back USB Computer Pass/Fail Record Type edge Source Slope Rising TRIGGER (edge) Mode Auto Coupling LEVEL 0.00V Type pulse Source When TRIGGER (pulse)
  • Page 134: Appendix C: Daily Maintain And Cleaning

    Appendix C: Daily Maintain and Cleaning Daily Maintain Do not store or leave the instrument in where the LCD display will be exposed to direct sunlight for long periods of time. CAUTION: To avoid damage to the instrument or probes, do not expose them to sprays, liquids, or solvents Cleaning If this instrument requires cleaning, disconnect it from all power sources...