SDS1000X-E&SDS1000X-U User Manual
You can also customize the probe attenuation factor. Press the Probe softkey,
select Custom, and then press the Custom softkey. Set the desired probe
attenuation ratio by Universal Knob or virtual keyboard. The range is [1E-6,1E6].
Unit
Select the amplitude display unit for the current channel. The available units are V
and A. When the unit is changed, the unit displayed in the channel label will
change accordingly.
1.
Press CH1button on the front panel to enter the CH1 function menu.
2.
Press the Next Page softkey to enter the second page of the CH1 function
menu.
3.
Press the Unit softkey to select the desired unit V or A.
The default setup is V.
Deskew
The valid range of each analog channel is ±100ns.
1.
Press CH1button on the front panel to enter the CH1 function menu.
2.
Press the Next Page softkey to enter the second page of the CH1 function
menu.
3.
Press the Deskew softkey. Then turn the Universal Knob to change deskew.
Invert
When Invert is set to On, the voltage values of the displayed waveform are inverted.
Invert affects how a channel is displayed and it keeps the trigger settings.
Inverting a channel also changes the result of any math function selected and
measure function.
1.
Press CH1button on the front panel to enter the CH1 function menu.
2.
Press the Next Page softkey to enter the second page of the CH1 function
32
WWW.SIGLENT.COM