SDS1000X-E&SDS1000X-U User Manual

Peak Detect

In this mode, the oscilloscope acquires the maximum and minimum values of the
signal within the sample interval to get the envelope of the signal or the narrow
pulse of the signal that might be lost. In this mode, signal confusion can be
prevented but the noise displayed would be larger.
In this mode, the oscilloscope can display all the pulses with pulse widths at least
as wide as the sample period.
Figure 22 Pulse With 0.1% Duty, Normal Mode
50
WWW.SIGLENT.COM