SDS1000X-E&SDS1000X-U User Manual
Data Length — When SDA data's length is equal to the value of Byte Length
and address's length is the same as set value, the oscilloscope will be
triggered. Byte length is in the range of 1 to 12 bits.
Operation Steps
1.
Press the Setup to enter the TRIGGER function menu.
2.
Press the Type and select Serial.
3.
Press the Protocol and select I2C.
4.
Press the Trigger Setting softkey.
5.
Press the Condition softkey and turn the Universal Knob to select the trigger
condition:
If you select the EEPROM condition:
a.
Press the Limit Range softkey to set the qualifier (=, < or >).
b.
Press the Data1 softkey and set its value by turning the Universal
Knob.
If you select 7 Addr & Data or 10 Addr & Data condition:
a.
Press the Addr softkey and turn the Universal Knob to select the 7- bit
or 10- bit device address.
b.
Press the Data1 or Data2 softkey and set the value about them.
c.
Press the R/W bit softkey and select write-frame or read-frame to
Figure 43 I2C TRIGGER Menu