compensated probes may cause measurement inaccuracy or error. The probe
compensation procedures are as follows.
1.
Set the switch to 10X on the probe.
2.
Perform steps 1, 2, 3 and 4 of "Function Inspection" in the previous section.
3.
Check the waveforms displayed and compare them with the following:
Under
Compensated
4.
Use a nonmetallic driver to adjust the low-frequency compensation adjustment
hole on the probe until the waveform displayed is as the "Perfectly
compensated" in the figure above.
Perfectly
Compensated
8
Over
Compensated