Lin Trigger And Serial Decode; Setup For Lin Signals - SIGLENT SDS1000X-E Series User Manual

Digital oscilloscope
Hide thumbs Also See for SDS1000X-E Series:
Table of Contents

Advertisement

SDS1000X-E&SDS1000X-U User Manual

LIN Trigger and Serial Decode

Please take the order of "Setup for LIN Signals", "LIN Trigger" and "LIN Serial
Decode" to trigger and decode the signals.

Setup for LIN Signals

There are two steps of setting the LIN signal, connecting the signal to oscilloscope,
specifying the parameters of each input signal.
1.
Press the Decode button to enter the DECODE function menu.
2.
Press the Decode softkey and select the desired slot (Decode1 or Decode2).
3.
Press the Protocol softkey and then select LIN by turning Universal Knob.
4.
Press the Signal softkey to enter the SIGNAL menu as Figure 58 shows.
5.
Press the Source softkey to select the channel that is connected to the LIN
signal.
6.
Press the Threshold softkey and set the LIN signal's threshold voltage level by
Universal Knob. The threshold voltage level is for decoding, and it will be
regard as the trigger voltage level when set the trigger type to serial.
7.
Press the
8.
Press the Configure softkey to enter the BUS CONFIG menu.
9.
Press the Baud softkey to set baud rate.
The baud rate can be set as predefined value.
If the desired baud rate is not listed, select custom option, press the
Custom and turn the Universal Knob to set the desired baud rate.
Figure 64 LIN SIGNAL Menu
softkey to return previous menu.
115
WWW.SIGLENT.COM

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents