Equipment Required; Post-Repair Adjustments - Agilent Technologies 4418B Service Manual

Power meter
Table of Contents

Advertisement

Equipment Required

The adjustment procedures include a list of recommended test equipment.
The test equipment is also identified on the test setup diagrams.

Post-Repair Adjustments

Table 3-1 lists the adjustments related to repairs or replacement of any of
the assemblies.
Table 3-1: Post Repair Adjustments, Tests, and Checks
Assembly Replaced
A1 Power Supply
A2 Processor Assembly
A3 Front Panel Assembly
A4 Comms Assembly
A5 Daughter Assembly
A6 Measurement Assembly
W1 or W2 Power Sensor
Cable
BT1 Lithium Battery
Power Reference Cable Kits
Agilent E4418B/E4419B Service Guide
Related Adjustments, Performance
Tests or Self Tests
Self Test.
Frequency and Level Adjustments,
Power Reference Level Test, Display
Brightness and Contrast Adjustment.
Instrument Accuracy Test, Display
Brightness and Contrast Adjustment.
All automated Tests and Serial Interface
Self Test.
Instrument Accuracy Test.
Zero Test, Instrument Accuracy Test.
Instrument Accuracy Test, Zero Test.
Battery Self Test.
Power Reference Level Test.
Adjustments
Introduction
3-3