Installing the Dryers

Prior to operating the 3050-AM Analyzer, the dryers must be installed. Two
(2) dryers have been supplied with this analyzer. Make sure that you have the
two (2) new dryers, and four (4) new VCR gaskets available before beginning
installation.
To install the dryers (see Figures 2-4.1 and 2-4.2):
1. Loosen the four (4) captured screws that secure the Dryer Compartment
2. Install one (1) dryer in the top position (Sample System Dryer) and one (1)
3. Remove the dryer mounting screws that secure the dryers in place and set
4. Working with one dryer at a time to minimize the time that the device is
Dryer Compartment
Cover
Figure 2-4.1.
Dryer installation locations.
PN 305804902, Rev A
The amount of time that the dryer fittings are left open should be kept
to less than one (1) minute to avoid damage to the device.
Cover (on the analyzer's Rear Panel, see Figure 2-4.1), and remove the cover.
in the bottom position (Zero Dryer).
them aside.
exposed to room air, remove the first dryer from its packaging. Loosen the
1/8-inch VCR fittings.
Sample System Dryer
(top dryer)
Zero Dryer
(bottom dryer)
Dryer Mounting Screws
(2 each)
1/8" VCR Gasket
(2 ea dryer)
Installation and Start-Up
|
2-9