A23 IF Mux Revision Check

A23 IF Mux Revision Check
While the instrument is still powered on check to see what version of the A23 IF Mux is installed in
the instrument by pressing System, Show Hardware. Find the IF Mux-10GHz entry, as seen in
Figure
2, and record the Part # below.
Figure 2
Hardware Information Screen - A23 IF Mux
A23 IF Mux Part Number Installed: ___________________
Use the part number of the assembly to determine which configuration the instrument has:
If the A23 IF Mux is part number N9041-60163 or N9041-60193, the instrument is
Configuration A.
If the A23 IF Mux is any other part number, the instrument is Configuration B.
The instrument configuration will be used when it comes to installing the parts needed to connect
the 4 GHz bandwidth output of the A23 IF Mux to the input of the new A22 Ultra-Wideband Analog
IF assembly.