2. GETTING STARTED

2.1 Power up

The Helipath Quick Action Laboratory Stand will go through a Power Up sequence when the power switch is turned
on. The touch screen will display a Splash screen for 5 seconds. The Splash screen is shown below (Figure 2. 1 ) and
includes the model number, the firmware version number, and the serial number.
Figure 2.1
The Splash screen information can also be accessed through the Settings Menu, using the Settings Button.
(see Section 2.4)

2.2 Start Up Screen

Upon powering up, the Helipath Quick Action Lab stand will need to perform a Homing operation prior to using the stand.
The screen below will be displayed.
Figure 2.2
The Helipath Quick Action Lab stand will move up automatically during the homing process.
The operator must press the Homing Button (refer section 2.3.1) to initiate the Homing. The operator may stop the
Homing process at any point by pressing the Stop Button (section 2.3.3) When complete, the Main Menu will be
displayed. An error 106 message prior to the homing process indicates the stand is out of level by more than 7.5
degrees. Level the stand before proceeding.
M-HPQA-001
H E L I PAT H Q U I C K A C T I O N L A B O R AT O R Y S TA N D - O P E R AT I O N A L M A N U A L