Remove all other shipping caps immediately prior to connection of gas
supplies.
Connect Sample / Carrier Gas
The analyzer uses purified sample gas as a chromatographic carrier and
actuator gas. Matching the carrier gas to the sample will prevent upsets
caused by the elution of the extraneous balance peak. The instrument
connection is made at the "CARRIER / SAMPLE INLET" 1/4" Male VCR
port on rear panel. Pressurize external line to 80 psig (Minimum 70 psig,
Maximum 90 psig).
Connect Actuator Gas
This analyzer needs high purity inert gas or CDA as valve actuator gas.
See gas specifications for Actuator Gas in Specifications for details.
The instrument connection is made at the "Actuator Gas" 1/8" VICI port
on rear panel. Pressurize external line to 80 psig (Minimum 70 psig, Maxi-
mum 150 psig).
Rear View showing gases and power connections
User Manual | 29