Adjusting The Gas Cylinder Holder And Connecting The Oxygen; Cylinder - Dräger Babyleo TN500 Instructions For Use Manual

Hide thumbs Also See for Babyleo TN500:
Table of Contents

Advertisement

Assembly and preparation
5.2.2
Connecting the nurse call
WARNING
Risk of electric shock
If the interface connectors and the patient are touched at the same time, there is a
risk of an electric shock.
► Do not simultaneously touch the nurse call connector and the patient.
CAUTION
No alarm in event of a nurse call fault
The nurse call may fail in the event of an incorrect device configuration, operating
errors (e.g., nurse call connector incorrectly connected) or a fault in the connection
between the nurse call and the central hospital alarm system.Alarms are not
transmitted.
► Be aware that only high- and medium-priority alarms trigger the nurse call.
► Bear in mind that using the nurse call does not absolutely guarantee that alarms
will be transmitted.
► Take note of alarms directly on the device.
Procedure:
1 Push the nurse call connector into the nurse call socket so that it engages
audibly.
5.3
Adjusting the gas cylinder holder and connecting the
oxygen cylinder
This chapter describes how the gas cylinder holder is matched to the size of the
oxygen cylinder. It also describes how to connect the oxygen cylinder to the device.
Prerequisites:
– The gas cylinder holder is installed on the correct side of the device.
– A full oxygen cylinder is available.
– A pressure reducer has been screwed onto the oxygen cylinder valve.
– The following parts are available:
– Compressed gas hose
– 2 guard plates for the duct rails
– 3 mm hex key
58
Instructions for use Babyleo TN500 SW 1.0n