Safety During Cleaning And Service - Dräger Babyleo TN500 Instructions For Use Manual

Hide thumbs Also See for Babyleo TN500:
Table of Contents

Advertisement

Safety-related information
2.15

Safety during cleaning and service

2.15.1
Risk of infection
The device must be reprocessed, otherwise there is an increased risk of infection
and the function of the device may be impaired.
► Disinfect and clean the device thoroughly before using it for the first time, and
then once a week and every time the patient changes.
► Follow hospital hygiene regulations, including reprocessing intervals and
reprocessing procedures.
► Use validated reprocessing procedures.
► Reprocess reusable products after every use.
► Follow the manufacturer's instructions in respect of cleaning agents and
disinfectants.
► Reprocess the product as described in ''Reprocessing'' before carrying out
service work or returning it for repair.
2.15.2
Disinfectants and cleaning agents
Residues of disinfectants or cleaning agents left in the device after reprocessing
may put the patient at risk due to irritation to the skin or mucous membranes.
► Observe the instructions relating to disinfectants and cleaning agents.
► Make sure that the device is adequately ventilated after reprocessing.
2.15.3
Service
This product must be serviced at regular intervals and in the appropriate manner.
Failure to perform service measures correctly may result in personal injury and
property damage.
► Perform service work as described in ''Service''.
► Service work must only be performed by the target groups authorized to perform
the work in question.
► Only carry out service measures when there is no patient in the device.
► Perform calibration, adjustment, and verification of the scale (option) at the
specified intervals.
2.15.4
Undetected damage to the device
Reprocessing, wear and tear, and improper storage may damage the device. The
functional integrity of the device is no longer guaranteed. The patient is put at risk.
► Comply with the ambient conditions for operation and storage of the device.
► If a device is behaving suspiciously or is clearly malfunctioning, replace it with a
device in proper working order.
22