Reprocessing
178
7 Fit the mattress tray.
8 Place the heated mattress or the SoftBed
mattress on the mattress tray. If the heated
mattress and the scale are available, pull
the cables through the hose grommets in
the center of the sensor wall. Push the
cables into the cable guides and plug in the
plug on the rear of the sensor wall.
9 Insert the 2 mini-walls at both ends of the
mattress tray.
10 Insert the hose grommets in the following
positions:
– Sensor wall: 3 pieces, on each corner
and in the middle of the sensor wall
– Front access panel: 3 pieces, on each
corner and in the middle of the front
access panel
Check that the hose grommets are correctly
seated.
11 Fold the side access panels and the front
access panel upward. Make sure that all
access panels latch securely in place.
Instructions for use Babyleo TN500 SW 1.0n