Dräger Babyleo TN500 Instructions For Use Manual page 145

Hide thumbs Also See for Babyleo TN500:
Table of Contents

Advertisement

Vertical tab
Heading or parame-
ter
Skin mode
Tolerated deviation
for central skin
temp. ±
Advanced skin tem-
perature monitoring
Skin temperature
Kangaroo mode
Skin temperature
Tolerate cooling
Skin temperature
(option)
Instructions for use Babyleo TN500 SW 1.0n
Description
The skin temperature sensors
must be connected to the
device and attached to the
patient.
– Specify the upper and lower
alarm limit for the tolerated
deviation of the central skin
temperature.
When the Advanced skin
temperature monitoring func-
tion is activated, additional
alarm limits can be determined:
– Upper alarm limit for the dif-
ference between the central
and peripheral skin tem-
perature
– Lower alarm limit for the dif-
ference between the central
and peripheral skin tem-
perature
The skin temperature sensors
must be connected to the
device and attached to the
patient. Specify the following
alarm limits:
– The lower alarm limits for
the central and the periph-
eral skin temperature
– The upper alarm limit for
the difference between the
central and peripheral skin
temperature
– The lower alarm limit for the
difference between the cen-
tral and peripheral skin tem-
perature
Specify the following alarm lim-
its:
– The upper alarm limit for
the central skin temperature
– The lower alarm limit for the
central skin temperature
– The upper and lower alarm
limit for the peripheral skin
temperature
Configuration
Factory setting
0.6 °C (1.1 °F)
Off
5 °C (9 °F)
Off
35 °C (95 °F)
5 °C (9 °F)
Off
35 °C (95 °F)
30 °C (86 °F)
Off
145