Procedure:
1 Touch the Oxygen therapy control (1) in the therapy bar.
 The therapy control turns yellow.
Fig. 51
2 Set the required oxygen concentration using the rotary knob.
3 Press the rotary knob to confirm.
 The therapy control turns dark green.
✓ The oxygen concentration is set.
6.12.3
Adjusting the oxygen alarm limits
Prerequisite:
– The required oxygen concentration is set.
Procedure:
1 Touch the Alarms... button (1) in the main menu bar.
 The Alarm limits dialog page is displayed.
Fig. 52
98
1
Therapy control for oxygen
Alarm limits for oxygen
1
2
Instructions for use Babyleo TN500 SW 1.0n