Fig. 56
No. Designation
1
2
3
6.14.2
Starting kangaroo mode
This chapter describes how to start kangaroo mode.
Prerequisites:
– The device is in operation.
– The patient is inside the device.
– The mother or father lies in a comfortable position on a couch next to the device
or sits on a stool.
– The last settings in incubator operation have been checked:
– Air temperature
– Humidity
– Oxygen (option)
– Mattress core temperature (option)
These settings are adopted in kangaroo mode.
Instructions for use Babyleo TN500 SW 1.0n
1
Main screen with kangaroo mode activated
Kangaroo mode
Kangaroo
End kangaroo
2
3
Description
Starts kangaroo mode.
Shows the elapsed time since kanga-
roo mode began.
Ends kangaroo mode.
Operation