WARNING
Risk of patient injury during transfer
Device parts may be damaged during transfer. The device may tip over. The patient
may be injured and device functions may be impaired. Before transfer, perform the
following measures:
► Set the device to transfer height and observe the information in the chapter
"Transferring the patient".
► Make sure that there is always sufficient clearance above the radiant warmer
and that the radiant warmer is not damaged.
► Make sure that accessories, e.g., pressure reducers and cylinder valves, do not
protrude beyond the trolley during the transfer.
► When transferring the patient, the device must always be moved by two people.
► Avoid shocks and vibrations as much as possible, e.g., when crossing door
thresholds and passing through door frames.
Check the functional integrity of the accessories during and after the
intrahospital transport. Observe the chapter "Adjusting the device height".
Procedure:
1 Adjust the height of the device using the foot pedals. The correct transfer height
is reached when the red hatched label on the lifting column is completely
covered.
 The correct device height for transfer is set.
Fig. 80
2 Carefully fold additional devices on the mounted support arms, e.g., patient
monitors or syringe pumps, against the device.
3 Secure loose objects on mounted support arms to prevent them from falling
down.
4 Close the hood.
5 Remove the x-ray cassette from the x-ray tray. Slide the x-ray tray into the
device.
6 Slide the bed support into the device.
7 Slide the drawer into the device and lock the drawer.
Instructions for use Babyleo TN500 SW 1.0n