Troubleshooting
Alarm pri-
Alarm
ority
!! 163
Ambient temperature sen-
sor faulty
!
212
Audio paused key faulty
!
163
Bed tilt inaccurate
!
162
Bed-tilt mechanism faulty
!! 169
Centr. sensor missing,
heating off
!! 168
Centr. temp. sensor faulty,
heating off
!! 120
Central skin temp. sensor
faulty
!! 50
Central skin temp. sensor
missing
!! 211
Central skin temperature
high
!! 210
Central skin temperature
low
!! 219
Central skin temperature
too high
154
Cause
An internal fault occurred.
The audio paused key has
been pressed for more than
10 seconds.
An internal fault occurred.
An internal fault occurred.
The central skin temperature
sensor (yellow) is not con-
nected.
The central skin temperature
sensor (yellow) is faulty.
The central skin temperature
sensor (yellow) is faulty.
The central skin temperature
sensor (yellow) is not con-
nected.
The current central skin tem-
perature is above the set
alarm limit.
The current central skin tem-
perature is below the set
alarm limit.
The temperature sensor is
not correctly attached to the
patient.
The current central skin tem-
perature is above 39.5 °C
(103.1 °F).
Instructions for use Babyleo TN500 SW 1.0n
Remedy
Contact specialized service
personnel.
Release the audio paused
key.
Contact specialized service
personnel.
Contact specialized service
personnel.
Contact specialized service
personnel.
Make sure that the sensor is
correctly connected.
Replace the sensor.
Replace the sensor.
Make sure that the sensor is
correctly connected.
Check the settings.
Check the patient's condition.
Make sure that the sensor is
correctly attached to the
patient.
Remove external heat
sources.
Check the settings.
Check the patient's condition.
Make sure that the sensor is
correctly attached to the
patient.
Check the settings.
Check the patient's condition.
Make sure that the sensor is
correctly attached to the
patient.
Remove external heat
sources.