8 Check the accuracy of the air temperature regulation:
Set value
Accuracy
9 Check the accuracy of the skin temperature regulation:
Set value
Accuracy
10 Check the accuracy of the O
Set value
Accuracy
11 Check the accuracy of the humidity regulation and humidity measurement:
Set value
Accuracy
12 Check the operating parameters for the lifting column under Service menu >
Operating Data > Lifting Column. The lifting column must be replaced if one of
the two operating parameters exceeds the maximum value. Contact
DrägerService.
Operating parameter
Operating parameter 1
Operating parameter 2
12.2.4
Metrological checks
If required by applicable regulations, the following measurement functions must be
checked in the specified intervals:
12.2.4.1
Skin temperature measurement
Prerequisites:
– The display range for the skin temperature sensors is active.
– Simulation resistors are connected instead of the skin temperature sensors. The
simulation resistors must maintain the following values:
Resistance
1586 Ω
1412 Ω
1312 Ω
Instructions for use Babyleo TN500 SW 1.0n
36.0 °C (96.8 °F)
±0.8 °C (±1.44 °F)
36.0 °C (96.8 °F)
± 0.1 °C (± 0.18 °F)
regulation:
2
50 Vol%
±3 Vol%
70 %
±10 %
Maximum value
7680
576
Temperature
33.2 °C (91.8 °F)
36.0 °C (96.8 °F)
37.8 °C (100.0 °F)