Procedure:
1 Open the flap for the x-ray tray on one side of the device. Pull the x-ray tray (1)
out.
Fig. 70
2 Position the x-ray cassette within the markings and push the x-ray tray into the
device.
3 Close the flap in front of the x-ray tray.
NOTICE
Risk of impact with x-ray tube assembly
If the clearance above the device is too small, the x-ray tube assembly may hit
the hood of the device.
► Before aligning the x-ray tube assembly, make sure that there is enough
4 Position the x-ray tube assembly between the radiant warmer and the patient
compartment.
5 Use the light field of the x-ray device to set the aperture to determine the desired
image section.
6 X-ray the patient.
7 Take the image receiver out of the x-ray tray and read it.
✓ The patient's x-ray is available.