Additional information:
– According to the data from this study, the recommended air temperature for a
patient aged 1 to 7 days does not depend on the patient's weight. Consequently,
the weight range cannot be selected for patients of this age.
– The last measured weight is used for determining the recommended air
temperature range. If there is no scale in the device, the weight range must be
selected manually.
– The range for the recommended air temperature in the System setup dialog
window is automatically adjusted in line with the age. The required air
temperature must be set on the main screen.
– If the ClearView on setting or the Auto setting of the Humidity therapy control
is changed or the patient is clothed or covered, this must be taken into account
in the air temperature settings.
– The recommended air temperature applies for patients up to an age of 36 days
and < 2500 g (5.5 lbs).
For further information, see the following chapter: ''Starting air temperature
mode'' (page 87)
8.2

Transferring data to another device

This chapter describes how configuration data and patient data are transferred from
one device to another.
8.2.1
Overview of data export
The data can be exported with a USB mass storage device.
Examples:
– To transfer the configured default settings to several devices
– To transfer the patient data to a new device when a device is changed
The following table shows which data can be exported.
Button
Configuration
data
Patient data
The trend data and the logbook data cannot be exported.
134
Exported data
Settings in System setup... > Therapy, e.g., air temperature,
skin temperature, air humidity, mattress core temperature
Settings in System setup... > Screen layout, e.g., screen
brightness, screen lock
Settings in System setup... > Alarms, e.g., alarm limits for the
skin temperature
Settings in System setup... > Patient, e.g., the configured
patient data, such as name, date of birth, birth weight
Current settings, which have been determined using the ther-
apy controls, e.g., with the therapy control for air temperature,
skin temperature, air humidity, mattress core temperature
Current settings in Alarms
Instructions for use Babyleo TN500 SW 1.0n